29.12.2021 Views

Diagnostic ultrasound ( PDFDrive )

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

e24

APPENDIX

Ultrasound Artifacts: A Virtual Chapter

DOPPLER IMAGING ARTIFACTS

he appearance of color Doppler signal is afected by several

variables including color-write priority, gray-scale gain, and pulse

repetition frequency. If the color gain is too low, low might not

be visualized. If the gain is too high, artifactual low may be seen

in adjacent sot tissues, and thrombus within the vessel might

be missed.

Loss or Distortion of Doppler Information

Incorrect gain, wall-ilter, and velocity scale can all lead to loss

or distortion of Doppler signal. A rule of thumb for gain adjustment

in Doppler imaging is to turn up the color gain until noise

is encountered and then back of just until the noise clears from

the image. In estimating frequency, the rule of thumb is to use

a Doppler angle less than 60 degrees (but not 0). Wall ilters

eliminate low-frequency noise, but a high setting can lead to

loss of signal. In general, wall ilters should be kept at the lowest

practical level, typically in the range of 50 to 100 Hz.

θ = 0°

cos θ = 1.0

∆F = 1.0

θ = 60°

cos θ = 0.5

∆F = 0.5

θ = 90°

cos θ = 0.0

∆F = 0.0

FIG. A.44 Effect of Doppler Angle on Frequency Shift. At an angle

of 60 degrees, the detected frequency shift detected by the transducer

is only 50% of the shift detected at an angle of 0 degrees. At 90 degrees,

there is no relative movement of the target toward or away from the

transducer, and no frequency shift is detected. The detected Doppler

frequency shift is reduced in proportion to the cosine of the Doppler

angle. Because the cosine of the angle changes rapidly at angles above

60 degrees, the use of Doppler angles of less than 60 degrees is recommended

in making velocity estimates. 2 (See Chapter 1, Fig. 1.35.)

A (PRF = 700 Hz)

B (PRF = 4500 Hz)

FIG. A.45 Artifactual Lack of Flow. Color Doppler image of a carotid artery and jugular vein. In (A), the pulse repetition frequency (PRF) is

700 Hz and there is aliasing in the carotid artery, but slow low in the jugular vein is seen. In (B), the PRF is 4500 Hz, eliminating aliasing in the

artery but also suppressing the display of the low Doppler frequencies in the internal jugular vein. 2 (See Chapter 1, Fig. 1.43.)

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!