29.12.2021 Views

Diagnostic ultrasound ( PDFDrive )

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

CHAPTER 1 Physics of Ultrasound 21

A

B

FIG. 1.30 Multipath Artifact. (A) Mirror image of the uterus is created by relection of sound from an interface produced by gas in the rectum.

(B) Echoes relected from the wall of an ovarian cyst create complex echo paths that delay return of echoes to the transducer. In both examples,

the longer path of the relected sound results in the display of echoes at a greater depth than they should normally appear. In (A) this results in

an artifactual image of the uterus appearing in the location of the rectum. In (B) the effect is more subtle and more likely to cause misdiagnosis

because the artifact suggests a mural nodule in what is actually a simple ovarian cyst.

the beam will cause loss of clinically important information from

deep, low-amplitude relectors and small targets. Ultrasound

artifacts may alter the size, shape, and position of structures.

For example, a multipath artifact is created when the path of

the returning echo is not the one expected, resulting in display

of the echo at an improper location in the image (Fig. 1.30).

Shadowing and Enhancement

Although most artifacts degrade the ultrasound image and impede

interpretation, two artifacts of clinical value are shadowing and

enhancement. Again, shadowing results when an object (e.g.,

calculus) attenuates sound more rapidly than surrounding tissues.

Enhancement occurs when an object (e.g., cyst) attenuates less

than surrounding tissues. Failure of TGC applied to normal tissue

to compensate properly for the attenuation of more highly

attenuating (shadowing) or poorly attenuating (enhancing)

structures produces the artifact (Fig. 1.31). Because attenuation

increases with frequency, the efects of shadowing and enhancement

are greater at higher than at lower frequencies. he conspicuity

of shadowing and enhancement is reduced by excessive beam

width, improper focal zone placement, and use of spatial

compounding.

DOPPLER SONOGRAPHY

Conventional B-mode ultrasound imaging uses pulse-echo

transmission, detection, and display techniques. Brief pulses of

ultrasound energy emitted by the transducer are relected from

acoustic interfaces within the body. Precise timing allows

determination of the depth from which the echo originates. When

pulsed wave ultrasound is relected from an interface, the

backscattered (relected) signal contains amplitude, phase, and

frequency information (Fig. 1.32). his information permits

inference of the position, nature, and motion of the interface

relecting the pulse. B-mode ultrasound imaging uses only the

amplitude information in the backscattered signal to generate

the image, with diferences in the strength of relectors displayed

in the image in varying shades of gray. Rapidly moving targets,

such as red cells in the bloodstream, produce echoes of low

amplitude that are not usually displayed, resulting in a relatively

anechoic pattern within the lumens of large vessels.

Although gray-scale display relies on the amplitude of the

backscattered ultrasound signal, additional information is present

in the returning echoes that can be used to evaluate the motion

of moving targets. 16 When high-frequency sound impinges on

a stationary interface, the relected ultrasound has essentially

the same frequency or wavelength as the transmitted sound. If

the relecting interface is moving with respect to the sound beam

emitted from the transducer, however, there is a change in the

frequency of the sound scattered by the moving object (Fig.

1.33). his change in frequency is directly proportional to the

velocity of the relecting interface relative to the transducer and

is a result of the Doppler efect. he relationship of the returning

ultrasound frequency to the velocity of the relector is described

by the Doppler equation, as follows:

∆F = ( F − F ) = 2⋅F ⋅ v c

R T T

he Doppler frequency shit is ΔF; F R is the frequency of sound

relected from the moving target; F T is the frequency of sound

emitted from the transducer; v is the velocity of the target toward

the transducer; and c is the velocity of sound in the medium.

he Doppler frequency shit (ΔF) applies only if the target is

moving directly toward or away from the transducer (Fig. 1.34A).

In most clinical settings the direction of the ultrasound beam

is seldom directly toward or away from the direction of low,

and the ultrasound beam usually approaches the moving target

at an angle designated as the Doppler angle (Fig. 1.34B). In this

case, ΔF is reduced in proportion to the cosine of this angle, as

follows:

∆F = ( F − F ) = 2⋅F ⋅ v ⋅cosθ

c

R T T

where θ is the angle between the axis of low and the incident

ultrasound beam. If the Doppler angle can be measured, estimation

of low velocity is possible. Accurate estimation of target

velocity requires precise measurement of both the Doppler

frequency shit and the angle of insonation to the direction of

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!