CHAPTER 56 Pediatric Interventional Sonography 1949

Withdraw

Transducer

cable

Toward

Toward

Ultrasound monitor view

A

B

Think: Do I need to move hub of needle

toward or away from transducer

cable?

Insert

C

D

Toward

cable

E

FIG. 56.7 Changing Needle Angulation. (A) If the needle,

once localized, is not at the correct angle, the needle should be

partly withdrawn. (B) The hub is moved in relation to the position

of the transducer cable. (C) The needle is reinserted at the correct

angle. (D) and (E) The principle of moving the hub in relation to the

transducer cable applies in any position of the transducer-needle

combination.

Initial Puncture Device

Most drainage catheters need to be placed over at least a

0.035-inch-diameter guidewire. An initial puncture with a smallcaliber

needle and placement of a 0.018-inch wire requires an

exchange maneuver, and each passage of a wire or dilator adds

to the time and complexity of the procedure, increasing the risk

of losing position or kinking the wire, especially with most

0.018-inch wires, which are not robust.

Various prepackaged exchange sets are available commercially.

Some sets (i.e., micropuncture, Aprima Access set) have a dilator/

sheath that is placed over the 0.018-inch wire and then the dilator

is removed, leaving the larger sheath in place, which allows

placement of the 0.035-inch wire. hese systems are used for

some biliary procedures or where vital structures are very close

to the target.

he Yueh needle (Cook, Bloomington, IN) is a needle/sheath

device with small holes drilled in the end of the catheter, making