29.12.2021 Views

Diagnostic ultrasound ( PDFDrive )

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

CHAPTER 1 Physics of Ultrasound 9

FIG. 1.11 Dynamic Range. The ultrasound receiver must compress the wide range of amplitudes returning to the transducer into a range that

can be displayed to the user. Here, compression and remapping of the data to display dynamic ranges of 35, 40, 50, and 60 dB are shown. The

widest dynamic range shown (60 dB) permits the best differentiation of subtle differences in echo intensity and is preferred for most imaging

applications. The narrower ranges increase conspicuity of larger echo differences.

for some means of automatic TGC, the manual adjustment of

this control is one of the most important user controls and may

have a profound efect on the quality of the ultrasound image

provided for interpretation.

Another important function of the receiver is the compression

of the wide range of amplitudes returning to the transducer into

a range that can be displayed to the user. he ratio of the highest

to the lowest amplitudes that can be displayed may be expressed

in decibels and is referred to as the dynamic range. In a typical

clinical application, the range of relected signals may vary by a

factor of as much as 1 : 10 12 , resulting in a dynamic range of up

to 120 dB. Although the ampliiers used in ultrasound machines

are capable of handling this range of voltages, gray-scale displays

are limited to display a signal intensity range of only 35 to 40 dB.

Compression and remapping of the data are required to adapt

the dynamic range of the backscattered signal intensity to the

dynamic range of the display (Fig. 1.11). Compression is performed

in the receiver by selective ampliication of weaker signals.

Additional manual postprocessing controls permit the user to

map selectively the returning signal to the display. hese controls

afect the brightness of diferent echo levels in the image and

therefore determine the image contrast.

Image Display

Ultrasound signals may be displayed in several ways. Over the

years, imaging has evolved from simple A-mode (amplitudemode)

and bistable display to high-resolution, real-time, grayscale

imaging. he earliest A-mode devices displayed the voltage

produced across the transducer by the backscattered echo as a

vertical delection on the face of an oscilloscope. he horizontal

time sweep of the oscilloscope was calibrated to indicate the

distance from the transducer to the relecting surface. In this

form of display, the strength or amplitude of the relected sound

is indicated by the height of the vertical delection displayed on

the oscilloscope. With A-mode ultrasound, only the position

and strength of a relecting structure are recorded.

Another simple form of imaging, M-mode (motion-mode)

ultrasound, displays echo amplitude and shows the position of

A

B

A

B

FIG. 1.12 M-Mode Display. M-mode ultrasound displays changes

of echo amplitude and position with time. Display of changes in echo

position is useful in the evaluation of rapidly moving structures such as

cardiac valves and chamber walls. Here, the three major moving structures

in the upper gray-scale image of the fetus are recorded in the corresponding

M-mode image and include the near ventricular wall (A), the interventricular

septum (B), and the far ventricular wall (C). The baseline is

a time scale that permits the calculation of heart rate from the M-mode

data.

moving relectors (Fig. 1.12). M-mode imaging uses the brightness

of the display to indicate the intensity of the relected signal.

he time base of the display can be adjusted to allow for varying

degrees of temporal resolution, as dictated by clinical application.

M-mode ultrasound is interpreted by assessing motion patterns

of speciic relectors and determining anatomic relationships

from characteristic patterns of motion. Currently, the major

application of M-mode display is evaluation of embryonic and

fetal heart rates, as well as in echocardiography, the rapid motion

of cardiac valves and of cardiac chamber and vessel walls. M-mode

imaging may play a future role in measurement of subtle changes

in vessel wall elasticity accompanying atherogenesis.

he mainstay of imaging with ultrasound is provided by

real-time, gray-scale, B-mode display, in which variations in

display intensity or brightness are used to indicate relected signals

C

C

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!