29.12.2021 Views

Diagnostic ultrasound ( PDFDrive )

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

32 PART I Physics

A

B

C

D

FIG. 1.46 Aliasing. Pulse repetition frequency (PRF) determines the sampling rate of a given Doppler frequency. (A) If PRF (arrows) is suficient,

the sampled waveform (orange curve) will accurately estimate the frequency being sampled (yellow curve). (B) If PRF is less than half the frequency

being measured, undersampling will result in a lower frequency shift being displayed (orange curve). (C) In a clinical setting, aliasing appears in

the spectral display as a “wraparound” of the higher frequencies to display below the baseline. (D) In color Doppler display, aliasing results in a

wraparound of the frequency color map from one low direction to the opposite direction, passing through a transition of unsaturated color. The

velocity throughout the vessel is constant, but aliasing appears only in portions of the vessel because of the effect of the Doppler angle on the

Doppler frequency shift. As the angle increases, the Doppler frequency shift decreases, and aliasing is no longer seen.

zone. his has led to an interest in HIFU as a means of destroying

noninvasive tumor and controlling bleeding and cardiac conduction

anomalies.

HIFU exploits thermal (heating of tissues) and mechanical

(cavitation) bioefect mechanisms. As ultrasound passes through

tissue, attenuation occurs through scattering and absorption.

Scattering of ultrasound results in the return of some of the

transmitted energy to the transducer, where it is detected and

used to produce an image, or Doppler display. he remaining

energy is transmitted to the molecules in the acoustic ield and

produces heating. At the spatial peak temporal average (SPTA),

intensities of 50 to 500 mW/cm 2 used for imaging and Doppler,

heating is minimal, and no observable bioefects related to tissue

heating in humans have yet been documented with clinical devices.

With higher intensities, however, tissue heating suicient to

destroy tissue may be achieved. Using HIFU at 1 to 3 MHz, focal

peak intensities of 5000 to 20,000 W/cm 2 may be achieved. his

energy can be delivered to a small point several millimeters in

size, producing rapid temperature elevation and resulting in tissue

coagulation, with little damage to adjacent tissues (Fig. 1.48).

he destruction of tissue is a function of the temperature reached

and the duration of the temperature elevation. In general, elevation

of tissue to a temperature of 60°C for 1 second is suicient to

produce coagulation necrosis.

Because of its ability to produce highly localized tissue destruction,

HIFU has been investigated as a tool for noninvasive or

minimally invasive treatment of bleeding sites, uterine ibroids,

and tumors in the prostate, liver, and breast. 19,20 As with diagnostic

ultrasound, HIFU is limited by the presence of gas or bone

interposed between the transducer and the target tissue. he

relection of high-energy ultrasound from strong interfaces

produced by bowel gas, aerated lung, or bone may result in tissue

heating along the relected path of the sound, producing unintended

tissue damage.

Major challenges with HIFU include image guidance and

accurate monitoring of therapy as it is being delivered. Magnetic

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!