Study of Indoor Air Quality in Two Different Laboratories at Kompleks<br />

Makmal Berpusat, UMT<br />

Noor Izatul Aekmal Binti Wan Azmi<br />

Supervisor: Dr. Nurul Adyani Binti Ghazali<br />

Bachelor of Technology (Environment)<br />

School of Ocean Engineering<br />

Universiti Malaysia Terengganu<br />

Indoor Air Quality (IAQ) in laboratories can be defined as the air quality inside a lab<br />

that will lead to the comfort and health of the occupants. The aim of this study was<br />

to measure the concentration of the temperature, relative humidity (RH) and<br />

Particulate Matter (PM10), in two different laboratories. The measurements were<br />

carried out during the weekdays for 3 days from 0830 to 1630 by using indoor air<br />

quality instruments. The results show the concentration of the PM10 and temperature<br />

are below the limit from Industry Code of Practice 2010 (ICOP 2010) with 33 µg/m 3<br />

and 25.8°C respectively while relative humidity exceed the limit with 71.1%. Results<br />

from ANOVA show that there is significant differences recorded for all parameters. As<br />

for conclusion, from the three parameters one is exceeded the limit therefore, this<br />

study need more concern to provide a better thermal comfort to peoples.<br />